Standard Dimension® QuikLYTE® Integrated Multisensor IMT Standard B 3 X 300 mL For Dimension Clinical Chemistry Systems

QUICK LIGHT, STD B D/S

No Image

Features

  • Standard B – used during the IMT 2-point calibration
  • The difference between Stnd A and Stnd B at each electrode is used to calculate the calibration slope
  • To determine if the calibration is successful, the system compares these slope values to the slope ranges stored in computer memory for each electrode
  • More …
  Log In to Order
or

Product Specifications


Features
  • Standard B – used during the IMT 2-point calibration
  • The difference between Stnd A and Stnd B at each electrode is used to calculate the calibration slope
  • To determine if the calibration is successful, the system compares these slope values to the slope ranges stored in computer memory for each electrode
  • Values assigned to the Dimension® QuikLYTE Standard B (S625) for the NA (Sodium) method are traceable to the National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 2201