Product Comparison


  SPS Medical Supply LCR-025 McKesson Brand 73-SBD030 Solventum Corporation 00135LF Solventum Corporation 1233LF Steris EQC009
McKesson # 1206253 524874 724429 724430 514538
Description Lantor Cube® Sterilization Bowie-Dick Test Pack Steam McKesson Sterilization Bowie-Dick Test Pack Steam Comply™ Sterilization Bowie-Dick Test Pack Steam Comply™ Sterilization Bowie-Dick Test Pack Steam Verify® Sterilization Bowie-Dick Test Pack Steam
Manufacturer # LCR-02573-SBD03000135LF1233LFEQC009
Brand Lantor Cube®McKessonComply™Comply™Verify®
Manufacturer SPS Medical SupplyMcKesson BrandSolventum CorporationSolventum CorporationSteris
Invoice TEST PACK, BOWIE DICK RUSBL (25/BX)TEST PACK, BOWIE-DICK DISP (30TESTS/CS)TEST KIT, BOWIE DICK LF (6PK/BG 5BG/CS) 3MTEST, BOWIE DICK TYPE LF (6/BG5BG/CS) 3MTEST KIT, VERIFY BOWIE DICK (20/BX)
    Log In to Order   Log In to Order   Log In to Order   Log In to Order   Log In to Order

Comparison

Country of Origin United StatesUnited StatesUnknownUnknownUnited States
Indicator Change Blue to BlackYellow to Brown-BlackYellow to Brown-BlackYellow to Blue-Purple
Is_ECAT N
Is_GSA N
Is_VA Y
Latex Free Indicator Not Made with Natural Rubber Latex
Temperature Range 273°F (134°C)270° to 273°F (132° to 134°C)270° to 273°F (132° to 134°C)270°F (132°C) / 273°F (134°C) / 275°F (135°C)
Time to Results 3.5 Minute Processing Time4 Minute Processing Time4 Minute Processing Time3.5 to 4 Minute Processing Time
Application Sterilization Bowie-Dick Test PackSterilization Bowie-Dick Test PackSterilization Bowie-Dick Test PackSterilization Bowie-Dick Test PackSterilization Bowie-Dick Test Pack
Class Class 2Class 2Class 2Class 2Class 2
Sterilization Compatibility SteamSteamSteamSteamSteam
UNSPSC Code 4228181042281810422818104228181042281810
Features and Benefits
  • 25 test sheets per pack and can be utilized 25 times before disposal (Requires use of metal clamp - one time purchase)
  • Reduces waste and inventory space
  • Easy to interpret indicator sheet with blue to dark color transition
  • Designed for simulating a 4 kg cotton pack according to ISO 11140-5 Type 2
  • McKesson Disposable Bowie-Dick Type Test Pack
  • Designed to detect the presence of residual air in pre-vacuum steam sterilizers.
  • For daily monitoring pre-vacuum steam sterilizers.
  • Each disposable test pack contains a chemical indicator sheet, and Instructions For Use are printed on the box.
  • Not made with natural rubber latex.
  • Packaged: 30 Test Packs Per Case
  • Early warning feature of the Comply Bowie-Dick Plus Test Pack permits scheduled sterilizer maintenance and helps to prevent unexpected failures
  • Lead Free ink chemistry
  • Early warning feature of the Comply Bowie-Dick Plus Test Pack permits scheduled sterilizer maintenance and helps to prevent unexpected failures
  • Easy-to-read test sheet clearly identifies failures and can be filed for record keeping
  • Equivalent to the Bowie-Dick Towel Pack cited in AAMI
  • Two year shelf life
  • Easy-to-read test sheet clearly identifies failures and can be filed for record keeping
  • Test pack will reproducibly detect residual air in the sterilizer chamber due to air leaks, air removal failures and inadequate vacuum by testing the air removal efficiency of 132-134˚C (270-273˚F) dynamic air-removal steam sterilizers
  • Indicator sheet will turn a uniform dark brown/black color except when air removal failures, such as air leaks, occur; can be filed for record keeping
  • Lead-free chemistry
  • Two year shelf life
  • Contains Latex
  • Consists of a series of steam penetration (air removal) barriers in the center of which is a chemical indicator sheet
  • The indicator sheet may be stored as a permanent record of the test.
  • Used to qualify newly installed prevacuum sterilizers or following a major repair
  • Designed to monitor performance changes in air removal and steam quality during routine use
  • Equivalent to AAMI Bowie-Dick Test Pack