Comparison |
Analyzer Type |
Benchtop | Floorstanding | Floorstanding | Benchtop | Floorstanding |
Automation Type |
| Automated | Automated | | Automated |
CLIA Classification |
CLIA Moderate Complexity | | CLIA Moderate Complexity | CLIA Moderate Complexity | |
Contents 1 |
| ISE Reagent Rental | Analyzer with ISE Module | | Without ISE Sales Group |
Country of Origin |
United States | Unknown | Unknown | Unknown | United States |
Dimensions |
23 X 27.5 X 22.5 Inch | 35 X 49 X 51 Inch | 760 X 1205 X 1450 mm | 40 X 28 X 25 Inch | 35 X 49 X 51 Inch |
Number of Tests |
40 Tests | 120 Tests | | 55 Assays | 120 Tests |
Power Source |
100 - 240 VAC, 20 Amp Circuit Required | 200 to 240 V, 50/60 Hz, 3.8 kVA | 210 V | | 200 to 240 V, 50/60 Hz, 3.8 kVA |
Purchase Program Type |
Placement Program | Rental Agreement | Standard Purchase Agreement | Standard Purchase Agreement | Standard Purchase Agreement |
Readout Type |
Printout | | | | |
Sample Capacity |
3 µL Min to 500 µL Max | 1 to 25 μL in 0.1 μL Increments | Up to 80 Sample Capacity | | 1 to 25 μL in 0.1 μL Increments |
Sample Type |
Whole Blood / Serum / Plasma / Urine Sample | Whole Blood (HbA1c) / Serum / Plasma / Urine / Other Fluids Sample | Serum / Plasma / Urine Sample | Whole Blood / Serum / Plasma / Urine / CSF / Homogenous Fluids Sample | Whole Blood (HbA1c) / Serum / Plasma / Urine / Other Fluids Sample |
Test Name |
| Lipemia, Hemolysis and Icterus Indices Clot Detection and Probe Crash Protection, 117 Photometrics (113 + LIH + HbA1c (Thb, HbA1c + HbA1c%)) and ISE | | | Lipemia, Hemolysis and Icterus Indices Clot Detection and Probe Crash Protection, 117 Photometrics (113 + LIH + HbA1c (Thb, HbA1c + HbA1c%)) and ISE |
Throughput |
165 Photometric Tests per Hour / 285 Tests per Hour with ISE | 800 Tests per Hour (Photometric) / Up to 1200 Tests per Hour (with ISE) / 200 Tests per Hour (ISE Sample) / 600 Tests per Hour (ISE Only, Maximum) | 400 Photometric Tests per Hour | 420 Tests per Hour Maximum (with ISE Module) | 800 Tests per Hour (Photometric) / Up to 1200 Tests per Hour (with ISE) / 200 Tests per Hour (ISE Sample) / 600 Tests per Hour (ISE Only, Maximum) |
Time to Results |
| Up to 8 Minutes 33 Second Results | | | Up to 8 Minutes 33 Second Results |
User Interface |
LCD Monitor | | | Touch Screen | |
Weight |
145 lbs. | 1015 lbs. | | 264 lbs. | 1015 lbs. |
Application |
Chemistry Analyzer with ISE | Chemistry Analyzer with ISE | Chemistry Analyzer with ISE | Chemistry Analyzer with ISE | Chemistry Analyzer with ISE |
CLIA Classified |
CLIA Non-Waived | CLIA Non-Waived | CLIA Non-Waived | CLIA Non-Waived | CLIA Non-Waived |
UNSPSC Code |
41115807 | 41115807 | 41115807 | 41115807 | 41115807 |
Features and Benefits |
- RAP placement program - The ACE Alera is a proven, high-performance, low-maintenance chemistry analyzer capable of improving patient care at a lower acquisition price
- The ACE Alera is best suited for 10-50 patients per day
- The ACE Alera system's patented closed tube sampling assures safety by effectively minimizing the risk of blood borne pathogens
- The ACE Alera’s refrigerated compartment maintains specimen and reagent integrity as well as eliminating overnight reagent transfer
- 40 bottle reagent compartment capacity (20 with 30 mL capacity: 20 with 12 mL capacity)
- 200 different assays programmed / calibrated at one time
- 40 different assays measured onboard simultaneously
- Batch and STAT test modes
- Assay types: final point, delta (two-point), slope (factor or calibrated) and quadratic
- Assay methods: photometry, potentiometry, turbidimetric/homogeneous EIA
- Operative methods: random access and continuous random access
- IBM-compatible processor
- Has a 40 GB hard drive
- Utilizes Okidata B4600 laser printer
- On screen error messages
- Utilizes Dot coded reagent bottles
- Holds up to 1,000 patient records
|
- The DxC 700 AU delivers increased uptime, high reliability and precise performance to mid- to large-sized laboratories and hospitals. Its flexible design offers standalone operation or connectivity to laboratory automation systems; with throughput of up to 800 photometric tests per hour (up to 1,200 with ISEs) and 63 onboard parameters, the DxC 700 AU delivers reliability and efficiency to laboratories around the world
- Fully automated, random-access clinical chemistry system with STAT capability
- Spectrophotometry and potentiometry analytical principles
- Colorimetry, turbidimetry, latex agglutination and indirect ISE analytical methods
- Endpoint, rate, fixed point and indirect ISE assay types
- Sampler capacity rack sampler: 10 samples per rack (barcodes on primary tubes and on racks), capacity for 150 samples, refrigerated STAT carousel: 22 samples can be run simultaneously (cal, QC and routine samples)
- Wavelength: 13 Different wavelengths between 340 nm and 800 nm
- 0 to 3 OD Photometric range
- Dry bath reaction method
- Permanent glass cuvettes reaction cuvette
- 150 Sample continuous rack loader
- Economical ISE's with long onboard stability; easy-to-maintain (only individual electrode replacement required)
- Priority sample repeat/reflex
- Ability to load reagents while the analyzer is in measure mode
- Cooled STAT compartment, providing one-button STAT interrupt, advanced auto QC and calibration capabilities
|
- The AU480 is the ideal primary chemistry analyzer for small- to medium-sized hospitals and laboratories, or as a dedicated specialty chemical or stat analyzer for larger institutions
- With random access throughput of up to 400 photometric tests per hour (up to 800 with electrolytes), increased onboard testing, reduced sample volume and user-friendly operation
- Cooled STAT module provides one button STAT interrupt, advanced Auto QC/calibration capabilities
- High-precision micro sampling
- Economical ISEs with long onboard stability
- 80-sample continuous rack loader
|
- The ABX Pentra C400 is an extremely compact Clinical Chemistry benchtop analyzer
- Includes optional ISE module (Na+, K+, Cl-_
- 55 on-board assays
- Innovative reagent management in cassette format
- Integrated work station and validation station
- Real-time viewing of your workflow
- Spectrophotometry: (Colorimetry and Turbidimetry) and Potentiometry: Direct (Serum or Plasma) and Indirect (Urine)
- 52 Reagent positions on board: 44 in a closed refrigerated area and 8 at room temperature
- FDA cleared
- Sample Capacity: 6 racks with 10 sample positions for primary tubes or cups
- Continuous loading
- Positive Sample Identification: Integrated Barcode Reader and Tube Presence Detector
- Sample Volume: 2 to 380 µL
- Calibrator/Controls Positioning on Sample or Reagent Tray
|
- The DxC 700 AU delivers increased uptime, high reliability and precise performance to mid- to large-sized laboratories and hospitals. Its flexible design offers standalone operation or connectivity to laboratory automation systems; with throughput of up to 800 photometric tests per hour (up to 1,200 with ISEs) and 63 onboard parameters, the DxC 700 AU delivers reliability and efficiency to laboratories around the world
- Fully automated, random-access clinical chemistry system with STAT capability
- Spectrophotometry and potentiometry analytical principles
- Colorimetry, turbidimetry, latex agglutination and indirect ISE analytical methods
- Endpoint, rate, fixed point and indirect ISE assay types
- Sampler capacity rack sampler: 10 samples per rack (barcodes on primary tubes and on racks), capacity for 150 samples, refrigerated STAT carousel: 22 samples can be run simultaneously (cal, QC and routine samples)
- Wavelength: 13 Different wavelengths between 340 nm and 800 nm
- 0 to 3 OD Photometric range
- Dry bath reaction method
- Permanent glass cuvettes reaction cuvette
- 150 Sample continuous rack loader
- Economical ISE's with long onboard stability; easy-to-maintain (only individual electrode replacement required)
- Priority sample repeat/reflex
- Ability to load reagents while the analyzer is in measure mode